Besonders empfindliches Messgerät für geringste Staubkonzentrationen, speziell in Müllverbrennungsanlagen
|
Features |
|
In-situ measurement directly in the flue gas flow |
|
Automatic system tests and correction of measured values |
|
Self-calibration in 4-h cycle |
|
Optics and electronics in a hermetically sealed unit - no smoke gas can enter device |
|
Maintenance-friendly, thanks to optimal purge air conduction in front of the heated optical end plates | |
Direct access to all parameters via the operator display | |
Automatic measuring range selection in accordance with 17. BImSchV. in relation 1:3:9. | |
Applications |
|
Fine dust measurements in processes | |
Waste incineration | |
Filter monitoring and emission-value measuring in waste incineration and power plants |
|
|
measurements |
dust concentration |
measuring ranges |
0–1 mg/m3 ... 0–30 mg/m3 1), |
measuring principle |
back scattering |
flue gas temperature |
above dew point up to 320°C, |
flue gas pressure |
-50 up to +20 hPa |
duct diameter |
0.3 up to 4 m |
ambient temperature |
-20 up to +50°C |
Protection |
IP65 |
measuring outputs |
2 x 0/4–20 mA/500 Ohm, |
digital outputs |
3 relay outputs, |
digital inputs |
1 potential free input |
accuracy |
<1% of measuring range |
detection limit |
<1% of measuring range |
reference point drift |
<0.2% of measuring range/month |
zero point drift |
<0.2% of measuring range/month |
supply voltage |
115 / 230 VAC, 50 / 60 Hz, 50 VA |
dimensions (H x W x D) |
measuring head 565 x 310 x 200 mm |
weight |
18 kg |
remarks |
1) after gravimetric calibration |
purge air supply |
|
purge air quantity |
approx. 80 m3/h |
supply voltage |
115/230 VAC, 50/60 Hz, |
dimensions (H x W x D) |
350 x 550 x 500 mm |
weight |
12 kg |
protection |
IP54 |
Extremely sensitive device for measuring smoke spot number
|
Features |
|
In-situ measurement directly in the flue gas flow |
|
Automatic system tests and correction of measured values |
|
Self-calibration in 4-h cycle |
|
Optics and electronics in a hermetically sealed unit - no smoke gas can enter device |
|
Maintenance-friendly, thanks to optimal purge air conduction in front of the heated optical end plates | |
Direct access to all parameters via the operator display | |
Applications |
|
Smoke spot number measurement in furnaces designed for light fuel oil |
|
|
measurements |
smoke spot number |
measuring ranges |
smoke spot number 1–3 (5) |
measuring principle |
back scattering |
flue gas temperature |
above dew point up to 320°C, |
flue gas pressure |
-50 up to +20 hPa |
duct diameter |
0.3 up to 4 m |
ambient temperature |
-20 up to +50°C |
Protection |
IP65 |
measuring outputs |
2 x 0/4–20 mA/500 Ohm, |
digital outputs |
3 relay outputs, |
digital inputs |
1 potential free input |
accuracy |
<1% of measuring range |
detection limit |
<1% of measuring range |
reference point drift |
<0.2% of measuring range/month |
zero point drift |
<0.2% of measuring range/month |
supply voltage |
115 / 230 VAC, 50 / 60 Hz, 50 VA |
dimensions (H x W x D) |
measuring head 565 x 310 x 200 mm |
weight |
18 kg |
remarks |
- |
purge air supply |
|
purge air quantity |
approx. 80 m3/h |
supply voltage |
115/230 VAC, 50/60 Hz, |
dimensions (H x W x D) |
350 x 550 x 500 mm |
weight |
12 kg |
protection |
IP54 |
Standard system for plants with concentrations |
Features |
|
In-situ measuring procedure, continuous measurement |
|
Semi-conductor source with long service life |
|
Super-wide band diode (SWBD), which provides more stable measuring results in comparison to devices with conventional LEDs |
|
Powerful microprocessor technology |
|
Measured value display on LC display in opacity, extinction or in mg/m3 | |
Automatic function tests with correction of measured values in relation to soiling | |
Optics and electronics in a hermetically sealed unit - no smoke gas can enter device | |
Easy adjustment without additional equipment | |
Low-maintenance thanks to optimal purge air conduction | |
Heated pane | |
Applications |
|
Plants in which the dust concentration quantity needs to be measured, e.g.: | |
Furnace plants with semi-anthracite coal, brown coal, fuel oil and combined heating |
|
Converter plants, asphalt mixing plants |
|
Plants for cement manufacture |
|
|
measurements |
opacity, extinction |
switchable measuring ranges |
opacity: 0–20% ... 0–100% |
measuring principle |
transmission |
flue gas temperature |
above dew point up to 250°C, |
flue gas pressure |
-50 up to +20 hPa, optional higher |
duct diameter |
1 up to12 m, optional up to18 m |
ambient temperature |
-20 up to +50°C, optional higher |
protection |
IP65, Ex optional |
measuring outputs |
2 x 0 / 4–20 mA / 500 Ohm, |
digital outputs |
6 relay outputs, programmable , |
digital inputs |
6 potential free inputs, programmable |
accuracy |
<1% of measuring range |
detection limit |
<2% of measuring range |
reference point drift |
<0.4% of measuring range/month |
zero point drift |
<0.4% of measuring range/month |
supply voltage |
95–264 VAC, 47–63 Hz, 30 VA |
dimensions (H x W x D) |
measuring head 363 x 185 x 398 mm |
weight |
17 kg |
purge air supply |
|
purge air quantity |
approx. 80 m3/h |
supply voltage |
115/230 VAC, 50/60 Hz, 0.37 /0.43 kW |
dimensions (H x W x D) |
350 x 550 x 500 mm |
weight |
12 kg |
protection |
IP54 |
The reliable system for monitoring dust emissions on small plants |
Features |
|
In-situ measurement directly in the flue gas flow |
|
Digital operation with microprocessor |
|
Adjustable limit value |
|
Contact switch outputs for pre-alarm and auto-power off |
|
Switchable measuring value integral 8 / 64 seconds | |
Applications |
|
Compact device for smaller plants |
|
Heating stations, power stations |
|
Boiler plants in industry, barracks, hospitals, schools |
|
Dust extraction and filter plants |
|
Control of processes in the chemical industry |
|
|
measurements |
opacity, optional extinction |
measuring ranges |
0–100% OP |
measuring principle |
transmission |
flue gastemperature |
above dew point up to 200°C, optional up to 500°C |
flue gas pressure |
-50 up to +20 hPa, optional higher |
duct diameter |
400–6250 mm |
ambient temperature |
-20 up to +50°C |
protection |
IP65 |
measuring outputs |
0 / 4–20 mA / 400 Ohm |
digital outputs |
2 relay outputs, |
digital inputs |
- |
accuracy |
<2% of measuring range |
detection limit |
<3% of measuring range |
reference point drift |
<0,6% of measuring range/month |
zero point drift |
<0.2% of measuring range/month |
supply voltage |
115 / 230 VAC, 50 / 60 Hz, 30 VA |
dimensions (H x W x D) |
measuring head 160 x 150 x 314 mm |
Weight |
7 kg |
purge air supply |
|
purge air quantity |
approx. 80 m3/h |
supply voltage |
115/230 VAC, 50/60 Hz, 0.37 /0.43 kW |
dimensions (H x W x D) |
350 x 550 x 500 mm |
Weight |
12 kg |
Protection |
IP54 |