A measuring device for the continuous monitoring of the smallest concentration of particles in the ambient air (fine dust)
|
Features |
|
C-14 method, no measurable decrease in activity |
|
Lowest radioactivity of all beta gauges, usable without licence, or disclosure |
|
Automatic zero correction |
|
Pre-calibrated, no site-specific calibration required |
|
Mass-flow controlled sample flow 1m3/h |
|
Extraction of a constant sample flow, irrespective of the ambient temperature |
|
Repeated collection on the same spot, collected particles available for heavy metal analysis |
|
RS-232 interface and analog output, status signals |
|
Applications |
|
Immissions measuring systems for monitoring fine dust |
|
Mobile immissions-measuring |
|
Dust measurement in health and safety applications |
|
Interior dust measurements |
|
Measurement and collection of dust particles for heavy metal analysis |
|
Long-term background studies in ambient dust concentration |
|
Dust measurement and collection at problem sites and repositories |
|
Dust measurement for secondary emission of repositories (e.g. coal) |
|
Dust measurement in supply air and exhaust ducts |
|
|
measurements |
dust concentration |
measuring ranges |
0–0.1 ... 0–10 mg/m3 |
measuring principle |
beta ray absorption |
ambient temperature |
0 up to +50°C |
protection |
IP20 |
measuring outputs |
2 x 0 / 4–20 mA / 500 Ohm |
digital outputs |
8 relay outputs, |
digital inputs |
3 potential free inputs |
accuracy |
<2% of measuring range |
detection limit |
<0,001 mg/Nm3 |
reference point drift |
<1% of measuring range/month |
zero point drift |
automatic zero point correction |
supply voltage |
230 VAC / 50 Hz, 110 V / 60 Hz, |
dimensions (H x W x D) |
320 x 450 x 500 mm, |
weight |
26 kg |
tube length |
standard 2 m |